Free sampleRedactedPrint-ready21 CFR Part 11

Sample Audit Evidence Packet

A redacted Q1 evidence package from a real AssetOS customer — Meridian BioSciences. Asset records, calibrations, work orders, complete audit trail, and deletion log. This is what AssetOS hands an FDA inspector.

Audit Evidence Packet — 21 CFR Part 11
FacilityMeridian BioSciences — Manufacturing Unit 4
Address200 Sidney Street, Cambridge, MA 02139
Audit period1 January 2026 – 31 March 2026 (Q1)
QA contactSarah Chen, QA Manager · s.chen@[redacted]
Export date2026-04-01 09:00:00 UTC
SystemAssetOS 3.2.1
Total records142 (4 assets · 4 calibrations · 5 work orders · 15 audit entries · 2 deletions · 6 users)
§ 1

Asset Register Extract

4 assets
Asset IDNameLocationCriticalityLast PMNext PM due
AST-0041Mettler Toledo XS205 Analytical BalanceLab A — Rm 204CRITICAL14 Jan 202614 Apr 2026
AST-0067Thermo Fisher 1300 Series A2 Biosafety CabinetLab B — Rm 218CRITICAL22 Jan 202622 Jul 2026
AST-0089Getinge GE 4610 Autoclave (Series 3)Sterilization — Rm 112ACRITICAL5 Jan 20265 Jul 2026
AST-0112Agilent 1260 Infinity II HPLC SystemQC Lab — Rm 301CRITICAL19 Feb 202619 Aug 2026
§ 2

Calibration Records

4 records
Cal refAssetDateTechnicianResultApproved byApproved at
CAL-2026-041-01AST-004114 Jan 2026j.rileyPASSt.bach14 Jan 2026 15:02
CAL-2026-067-01AST-006722 Jan 2026p.nairPASSt.bach22 Jan 2026 16:44
CAL-2026-089-01AST-00898 Feb 2026j.rileyPASSs.chen8 Feb 2026 11:30
CAL-2026-112-01AST-011219 Feb 2026p.nairPASSs.chen19 Feb 2026 14:55

4 of 4 calibrations passed. Next due: AST-0041 — 14 Apr 2026.

§ 3

Work Order History

5 work orders
WO refAssetTypeDescriptionOpenedClosedTechApproved byStatus
WO-2026-0089AST-0089PREVENTIVEAnnual PM — gasket replacement, door seal inspection, cycle validation3 Jan 20265 Jan 2026j.rileyt.bachAPPROVED
WO-2026-0091AST-0041CALIBRATIONQ1 calibration — 5-point linearity check per SOP-QC-01412 Jan 202614 Jan 2026p.nairt.bachAPPROVED
WO-2026-0094AST-0067PREVENTIVESemi-annual certification — airflow test, HEPA integrity, UV lamp check19 Jan 202622 Jan 2026j.rileyt.bachAPPROVED
WO-2026-0097AST-0089CORRECTIVEUnscheduled — chamber door pressure seal degraded. Replaced per SOP-MAINT-022.2 Feb 20262 Feb 2026j.rileyt.bachAPPROVED
WO-2026-0103AST-0112PREVENTIVEHPLC column swap + post-swap system suitability qualification17 Feb 202619 Feb 2026p.nairs.chenAPPROVED
§ 4

Audit Trail Extract

15 entries (sample)

Every field change, status transition, file upload, and sign-off is logged server-side. Log entries are immutable — they cannot be edited or removed by any user including administrators.

2026-01-05 09:14:22 UTCj.rileyWO-2026-0089 status IN_PROGRESS → COMPLETE
2026-01-05 09:14:23 UTCj.rileyWO-2026-0089 completion_notes [null] → "PM complete. Gasket set replaced (Part #GK-4610-09). All checks within spec."
2026-01-05 09:41:07 UTCt.bachWO-2026-0089 status COMPLETE → APPROVED (supervisor sign-off)
2026-01-14 14:33:18 UTCp.nairCAL-2026-041-01 result [null] → PASS (observed: 0.00003 g, tolerance: ±0.0001 g)
2026-01-14 14:33:18 UTCp.nairCAL-2026-041-01 status IN_PROGRESS → COMPLETE
2026-01-14 15:02:44 UTCt.bachCAL-2026-041-01 status COMPLETE → APPROVED (QA sign-off)
2026-02-02 07:58:04 UTCj.rileyWO-2026-0097 created type=CORRECTIVE asset=AST-0089 priority=HIGH
2026-02-02 08:11:39 UTCj.rileyWO-2026-0097 evidence_photo uploaded "door_seal_pre_replacement_20260202.jpg" (1.8 MB)
2026-02-02 08:44:12 UTCj.rileyWO-2026-0097 status IN_PROGRESS → COMPLETE
2026-02-08 10:22:11 UTCj.rileyAST-0089 location "Sterilization Rm 112" → "Sterilization Rm 112A"
2026-02-08 10:22:12 UTCj.rileyAST-0089 change_reason "Facility room renumbering per Facilities Notice FN-2026-003"
2026-02-19 14:55:01 UTCs.chenCAL-2026-112-01 status COMPLETE → APPROVED (QA sign-off)
2026-03-15 16:44:02 UTCp.nairWO-2026-0103 evidence_photo uploaded "hplc_column_post_swap_20260315.jpg" (2.4 MB)
2026-03-22 08:11:33 UTCd.okoyeUSER j.riley role TECHNICIAN → SENIOR_TECHNICIAN
2026-03-31 17:00:00 UTCsystemScheduled report Q1-2026-AUDIT-PACK generated 142 records exported by s.chen
§ 5

Deletion Log — Dual Approval

2 deletions

Record deletion requires a second authorized user to approve the request. Both the request and approval are permanently logged. Deleted records are retained in this log with full attribution.

DEL-2026-00128 Feb 2026
Record deletedWO-2026-0098 (draft, never assigned — created in error)
Requested byd.okoye
ReasonCreated in error during bulk import test. Never assigned or actioned.
Approved bys.chen
Approval timestamp2026-02-28 11:04:17 UTC
DEL-2026-00210 Mar 2026
Record deletedCAL-2026-041-DUP (duplicate calibration entry created by system sync error)
Requested byp.nair
ReasonDuplicate of CAL-2026-041-01 created by integration sync on 2026-01-14. Original record intact.
Approved byt.bach
Approval timestamp2026-03-10 14:22:08 UTC
§ 6

User Access Summary

6 accounts
User IDNameRoleLast activeMFA
s.chenSarah ChenQA Manager31 Mar 2026ENABLED
d.okoyeDavid OkoyeSystem Administrator31 Mar 2026ENABLED
t.bachThomas BachMaintenance Supervisor31 Mar 2026ENABLED
j.rileyJames RileySenior Technician29 Mar 2026ENABLED
p.nairPriya NairTechnician28 Mar 2026ENABLED
ext.auditorFDA Audit AccountAuditor (read-only)15 Mar 2026ENABLED

No shared accounts. All 6 users have unique IDs. Auditor account is read-only and time-boxed (expires 30 Apr 2026).